Influência da profundidade da inserção do parafuso no pedículo vertebral sobre o torque de inserção e a resistência ao arrancamento

Influence of the new depth of insertion of pedicle screw on insertion torque and pullout strength

Influencia de la profundidad de la inserción del tornillo en el pedículo vertebral sobre el torque de inserción y la resistencia al arrancamiento

Vânia Custódio Leite, Rodrigo Okubo, Antonio Carlos Shimano, Frank Kandziora, Helton LA Defino

Downloads: 0
Views: 7

Resumo

Objetivo: avaliar a influência da profundidade da inserção do parafuso do sistema de fixação vertebral e o diâmetro do orifício piloto sobre o seu torque de inserção e a resistência ao arrancamento. Métodos: foram utilizadas vértebras de suínos e parafusos USS de 5mm de diâmetro. Os parafusos foram inseridos nos pedículos das vértebras após a realização de orifício com brocas de 2,5; 3,8 e 4,5mm. Foi utilizado torquímetro Mackena® para a medida o torque de inserção. Foram formados seis grupos experimentais de acordo com o diâmetro do orifício piloto e a inserção do parafuso (total e somente da rosca). Cada grupo experimental era formado por dez corpos de prova. Grupo 1 – orifício de 2,5mm e inserção total do parafuso; Grupo 2 – orifício de 2,5mm e inserção somente da parte rosqueada; Grupo 3 – orifício de 3,8mm e inserção total; Grupo 4 – orifício de 3,8mm e inserção somente da parte rosqueada; Grupo 5 – orifício de 4,5 mm e inserção total; Grupo 6 – orifício de 4,5mm e inserção somente da parte rosqueada. Os ensaios de arrancamento foram realizados na Máquina Universal de Ensaio EMIC®. Resultados: a média da força máxima de arrancamento e torque de inserção nos diferentes grupos experimentais foram, respectivamente: Grupo 1: (1014,7 ± 153,3) N e (1,39 ± 0,21) N.m; Grupo 2: (909,9 ± 109,9) N e (0,87 ± 0,22) N.m; Grupo 3: (996,9 ± 107,4) N e (1,04 ± 0,19) N.m; Grupo 4: (947,7 ± 165,4) N e (0,57 ± 0,13)N.m; Grupo 5: (862,6 ± 141,2) N e (0,78 ± 0,19) N.m e Grupo 6: (669,6 ± 176,9) N e (0,34 ± 0,11) N.m. Conclusão: foi observada diferença estatística entre o torque de inserção dos parafusos, total ou parcialmente inseridos, considerando-se todos os diâmetros de perfuração utilizados. Foi observada diferença estatística entre a força de arrancamento dos parafusos, total e parcialmente inseridos, nos grupos com perfuração menor e maior que o diâmetro interno do parafuso.

Palavras-chave

Coluna vertebral; Parafusos ósseos; Biomecânica

Abstract

Objective: to assess the influence of the depth of pedicle screw insertion on the pullout strength considering different pilot hole diameters. Method: pig vertebrae and 5mm USS screws were used in the study. The screw were inserted into the lumbar and lower thoracic vertebral pedicle.bral pedicle. The pilot holes were drilled with 2.5, 3.8 and 4.5 mm drill and a Mackenna® torque meter was use to assess the screw insertional torque. The following six experimental groups with 10 test models each, were used: Group 1 – 2.5 mm pilot hole and total insertion of the screw; Group 2 – 2.5 mm pilot hole and screw partial insertion (only the threaded part of the screw inserted); Group 3 – 3.8 mm pilot hole and total screw insertion; Group 4 – 3.8 mm pilot hole and screw partial insertion B; Group 5 – 4.5 mm pilot hole and screw total insertonl; Group 6 – 4.5 mm pilot hole and partial insertion. The screw pullout tests were performed using an EMIC® Universal Testing Machine Data were analyzed by ANOVA and the Tukey-Kramer test, with the level of significance set at 5%. Results: the mean maximum pullout strengths and insertion torques were, respectively: Group 1: (1014.7±153.3) N and (1.39±0.21) N.m; Group 2: (909.9±109.9)N and (0.87±0.22) N.m; Group 3: (996.9±107.4) N and (1.04±0.19) N.m; Group 4: (947.7±165.4) N and (0.57±0.13) N.m; Group 5: (862.6±141.2) N and (0.78±0.19) N.m and Group 6: (669.6±176.9) N and (0.34±0.11) N.m. Conclusion: it was observed significant difference of insertional torque between the total and partial insertion of the screw. The screw pullout strength was significantly different between the totally and partially inserted screws in the groups with pilot hole diameter smaller or bigger than the internal diameter of the screw.

Keywords

Spine; Bone screws; Biomechanics

Resumen

Objetivo: evaluar la influencia de la profundidad de la inserción del tornillo del sistema de fijación vertebral y el diámetro del orificio piloto sobre su fuerza de inserción y la resistencia al arrancamiento. Métodos: fueron utilizadas vértebras de cerdos y tornillos USS de 5mm de diámetro. Los tornillos fueron inseridos en los pedículo de las vértebras después de la realización del orificio con brocas de 2.5; 3.8 y 4.5mm. Fue utilizado un torquímetro Mackena® para la medida del torque de inserción. Fueron hechos 6 grupos experimentales según el diámetro del orificio piloto y la inserción del tornillo (total y solamente de la rosca). Cada grupo experimental era formado por 10 cuerpos de prueba. Grupo 1 – orifico de 2.5mm e inserción total del tornillo; Grupo 2 – orifico de 2.5mm e inserción solamente de la parte rosqueada; Grupo 3 – orifico de 3.8 mm e inserción total; Grupo 4 – orifico de 3.8mm e inserción solamente de la parte rosqueada; Grupo 5 – orificio de 4.5 mm e inserción total; Grupo 6 – orifico de 4.5 mm e inserción solamente de la parte rosqueada. Los ensayos de arrancamiento fueron realizados en la Máquina Universal de Ensayo EMIC®. Resultados: la media de la fuerza máxima de arrancamiento y torque de inserción en los diferentes grupos experimentales fueron respectivamente: Grupo 1 – (1014.7±153.3)N y (1.39±0.21) N.m; Grupo 2 – (909.9±109.9)N y (0.87±0.22) N.m; Grupo 3 – (996.9±107.4) N y (1.04±0.19) N.m; Grupo 4 – (947.7±165.4) N y (0.57±0.13) N.m; Grupo 5 – (862.6±141.2) N y (0.78±0.19) N.m y Grupo 6 – (669.6±176.9) N y (0.34±0.11) N.m. Conclusión: fue observada diferencia estadística entre el torque de inserción de los tronillos totalmente o parcialmente inseridos y esa diferencia fue observada considerándose todos los diámetros de perforación utilizados. Fue observada diferencia estadística entre la fuerza de arrancamiento de los tornillos totalmente y parcialmente inseridos en los grupos con perforación menor y mayor que el diámetro interno del tornillo.

Palabras clave

Columna vertebral; Tornillos óseos; Biomecánica

Submetido em:
27/03/2007

Aceito em:
20/04/2008